Mts Model FlexTest SE Instrukcja Użytkownika Strona 194

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 264
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 193
Models FlexTest® IIm/GT/SE Controller Hardware
Interlock Jumper Plugs
FlexTest IIm Controller Connections
194
The interlock connectors require jumpers under the following conditions:
If you do not use the external hydraulic/program interlocks on any of the D-
15S connectors (J3A Sta 1, J3B Sta 2, J3C Sta 3, and J3D Sta 4). Use an
external jumper plug (PN 052-750-501) on each connector not used.
If you use the system digital I/O as a stand-alone unit. That is, if you do not
connect the System Digital I/O Module to a 497.xx Chassis, you must use an
external jumper plug (PN 050-774-701) on the J5 497 Intlk connector.
This shows how to connect an external switch or set of relay contacts to one of
the station interlocks:
J3A-J3D
External Switch
or
Relay Contact
External Device
+12 VDC
x9x11
x1x4
Opto-
Coupler
1
9
7
Przeglądanie stron 193
1 2 ... 189 190 191 192 193 194 195 196 197 198 199 ... 263 264

Komentarze do niniejszej Instrukcji

Brak uwag