709 Alignment System
Software Installation
Installation
56
Note If the FieldPoint has already been configured, you can skip the next step.
6. If this is a new installation and the FieldPoint has not been configured
before, you will get a warning message saying values do not match current
file settings for each unconfigured module.
A. Click OK
B. Ensure FP-SG-140@1 is selected.
C. In the File/Device Conflict Resolution, select File and click Use
These Settings.
D. In the Channel Attributes, ensure the Excitation Voltage is 5
Volts.
E. Repeat for FP-SG-140@2 and FP-AI-110@3.
B
A
C
D
E