Strona 1 - Testing Software

MTS TestLine Test System OperationUsing MTS FlexTest Controller Software and MTS TestSuiteTesting Software100-275-890 A be certain.

Strona 2 - Proprietary Software

• Record the name of the person who helped you.• Write down any specific instructions.After you callMTS logs and tracks all calls to ensure that you r

Strona 3 - Table of Contents

9. Turn off the Function Generator.Click the Program Stop button to turn off the Function Generator.Check and Reset Meter HistoryCheck meter history a

Strona 4

Note: Acceptable limits are dependent on the specimen you are using for the test. For detailedinformation, consult your lab experts or a materials han

Strona 5

For this example choose limits from -0.6 kN to 1.0 kN.a) If necessary, click the Station Setup button.b) Select Channels > Ch1 > Force > Chan

Strona 6

2. Minimize the Station Manager application by clicking the Minimize button in the upper right corner.3. Verify that the Station Manager application i

Strona 7 - Technical Support

Select All Programs > MTS TestSuite > Multipurpose Express.2. Alternatively, you can open the MPX application by clicking on a desktop icon.The

Strona 8

Select the TestFrom the TestSuite main window, select Tests > Example HCF Force Test to load the test.If Necessary, Correct Any Resource Errors1. F

Strona 9 - If You Contact MTS by Phone

In this example, MPX had the axial channel named "Axial" while the controller had the axial channelnamed "Ch 1." By changing the M

Strona 10

3. Enter the test parameters as shown in the following window.4. Review the status information.Review the status information that appears. If necessar

Strona 11 - Documentation Conventions

Run the Test1. Run the test.a) If necessary, click the Reset button to reset the interlock.b) Ensure that both the HPU and HSM are on high power.c) Cl

Strona 12 - Caution:

Generate a Report1. From the MPX application Explorer panel, select Example HCF Force Test > Test Runs > Test Run8, and then right-click and sel

Strona 13

PrefaceBefore You BeginSafety first!Before you use your MTS product or system, read and understand the safety information provided withyour system. Im

Strona 14

Save the Test and Minimize MPXThe Station Manager application is used to remove the specimen, so the test can be saved and the MPXapplication minimize

Strona 15

Click the Minimize button in the upper right corner of the MPX application.Remove the Specimen1. Turn off the HSM.Set HSM power to low, and then to of

Strona 16 - General Safety Practices

3. Loosen the specimen angle bracket from the test bed fastener.4. Remove the angle bracket from the specimen fasteners.112 | MTS TestLine Test System

Strona 17

5. Remove the specimen from the load cell.If Necessary, Recover from a TrippedDisplacement LimitIf a displacement limit is exceeded, use the following

Strona 18

a) In the Station Manager application main window, click the Manual Command button.b) Ensure that Exclusive Control is checked.c) In the Manual Comman

Strona 19

b) Check the current upper and lower limit settings to identify the acceptable range for the actuator.Adjust the limits if incorrect. If correct, cont

Strona 20

In the Station Manager application main window, click the Manual Command button. Move the actuatorto within acceptable limits while the interlock is o

Strona 21

MaintenanceTopics:•Routine Maintenance Overview Checklist...118•Ma

Strona 22

Routine Maintenance Overview ChecklistRecommended service to be performed at each running time interval notedAnnuallyBiweeklyWeeklyDailyCalendar Time

Strona 23 - System Introduction

AnnuallyBiweeklyWeeklyDailyCalendar Time using 8 hour RunningTime rate per day2,0001,500100050080408Running Time-HoursMTSMTSMTSMTSCheck that manifold

Strona 24 - About Other MTS Documentation

Warning:Warning notices indicate the presence of a hazard with a medium level of risk which, if ignored,can result in death, severe personal injury, o

Strona 25 - System Overview

AnnuallyBiweeklyWeeklyDailyCalendar Time using 8 hour RunningTime rate per day2,0001,500100050080408Running Time-HoursMTSMTSMTSMTSCheck response to fu

Strona 26

When to Do ItWhat to DoAfter the valve balancing procedure (electricalcompensation) has been completed and the resultsare judged unsatisfactory.Perfor

Strona 28

DecommissioningTopics:•Decommission the System...

Strona 29 - Station Setup window

Decommission the SystemThe decommissioning process is performed when the system is going to be moved or taken out of service.Disassembly is required w

Strona 30

• When removing the hydraulic components, ensure that you have drain pans and towels availablethat are sufficient to handle the drain volume of the it

Strona 32

Index793 software, See MTS FlexTest (Series 793) controllersoftwareAactionsdetector 46actuatorhydraulic lines 56maintaining 118positioning 76signal li

Strona 33 - Application

force sensor (load cell) (continued)calibration verification 54formproblem submittal 10function generator 32Function Generatorrunning 69, 95turning of

Strona 34

Ooverviewtesting 50PP gaincorrect 73, 99excessive 72, 98insufficient 71, 97P Gainadjusting 69, 95parameter setidentifying 53parameter sets 41parameter

Strona 35

Electronic manual conventionsThis manual is available as an electronic document in the Portable Document File (PDF) format. It can beviewed on any com

Strona 36

test (continued)selectingMTS Multipurpose (MP) Express application 105status information 107test parameters 35test procedure files 42testing overview

Strona 38 - About This Chapter

MTS Systems Corporation14000 Technology DriveEden Prairie, Minnesota 55344-2290 USAToll Free Phone: 800-328-2255(within U.S. or Canada)Phone:952-937-4

Strona 40 - Understanding MTS File Types

SafetyTopics:•General Safety Practices...

Strona 41

General Safety PracticesIf you have system related responsibilities (that is, if you are an operator, service engineer, or maintenanceperson), you sho

Strona 42

Read all manualsStudy the contents of this manual and the other manuals provided with your system before attempting toperform any system function for

Strona 43

Know electrical hazardsWhen the system electrical power is turned on, minimize the potential for electrical shock hazards. Wearclothing and use tools

Strona 44

• Wear appropriate safety devices to protect your hearing. Escaping air or gas can create a noise levelthat can damage your hearing.• Ensure that all

Strona 45

©2014 MTS Systems Corporation. All rights reserved.Original Instructions (English): 100-275-890 A—May 2014—MTS TestSuite Multipurpose 2.6 or later / 7

Strona 46 - Using the E-Stop Control

(see the System Documentation CD). Particles present in the hydraulic fluid can cause erratic or poorsystem response.Protect accumulators from moving

Strona 47

Safety Practices While Operating the SystemWear appropriate personal protectionWear eye protection when you work with high-pressure hydraulic fluid, h

Strona 48

Do not disturb sensorsDo not bump, wiggle, adjust, disconnect, or otherwise disturb a sensor (such as an accelerometer orextensometer) or its connecti

Strona 49

System IntroductionTopics:•About This Manual...

Strona 50 - Station Equipment Checkup

About This ManualThis manual is for operators of MTS TestLine systems. It describes the following for a typical MTS TestLinesystem:• Primary component

Strona 51

System OverviewStation ComponentsDescriptionNameItemThe high pressure hand valve allows you to controlpressure to the HSM. Ensure that the hand valve

Strona 52

DescriptionNameItemHSM contains the ports to provide hydraulicfiltration, the use of accumulators, and hydraulicsolenoid valves to control the hydraul

Strona 53 - Verify Calibration Files

DescriptionNameItemThe controller provides the necessary interfacebetween the PC and the load frame.Controller7The remote E-stop quickly turns off the

Strona 54

• Moving the actuator (for specimen installation)• Using the function generator• Setting tuning parametersThe Station Manager application windows asso

Strona 55

Setting LimitsSetting limits helps prevent specimen damage and injury. You can set limits using the Limits tab in theStation Setup window.MTS TestLine

Strona 56

Table of ContentsTechnical SupportHow to Get Technical Support...

Strona 57

Setting Up MetersMeters allow you to monitor displacement and forces. Using the following windows, you can set up multiplemeters and arrange them to y

Strona 58

Moving the ActuatorDuring setup, you must move the actuator so that you can clamp the specimen in the grip. The ManualCommand window allows you to mov

Strona 59

Using the Function GeneratorThe function generator allows you to input a signal to the specimen so that you can monitor the feedback.This is necessary

Strona 60 - Set Initial Tuning Parameters

Setting Tuning ParametersTuning parameters are found in the Station Setup window. You can enter tuning parameters on theAdjustments tab.MTS TestSuite

Strona 61

Correcting Resource ErrorsDepending on your station configuration, the resources in the test may not match the resources for thestation. Resource erro

Strona 62

Setting Test ParametersOnce the test is selected, MTS TestSuite MPX provides you an opportunity to set test parameters via theSetup Variables window.R

Strona 63 - Power Up the Pump and HSM

36 | MTS TestLine Test System OperationSystem Introduction

Strona 64

Key ConceptsTopics:•About This Chapter...

Strona 65

About This ChapterThis chapter, as well as the information in the Safety chapter, contains information you should know beforeyou attempt to run tests

Strona 66

• Offsetting the weight of fixtures so those values do not appear in test data• Adjusting gain to optimize system responseMTS TestSuite Test Software

Strona 67

Station Equipment Checkup...50Turn on the Contr

Strona 68

Switching Between Station Manager and MPX When Running TestsDescriptionItemStation Manager Main Window1Task Bar2Set Up Tests Using the Station Manager

Strona 69

Station File TypesDescriptionFile TypeA FlexTest project is a collection of files related to the station configuration.When you open a configuration,

Strona 70

DescriptionFile TypeTest Template files are command files created with Multipurpose Elite forperforming tests on Station Configurations. Test Template

Strona 71 - Insufficient P Gain

How MTS Files Work TogetherThe basic workflow is as follows:• You use the Station Manager application to open a Station Configuration.• You select a P

Strona 72 - Excessive P Gain

Closed Loop Control Step-by-StepThe following closed-loop diagram illustrates how the system operates when you use the Manual Controlpanel in the Stat

Strona 73

Understanding Control Channels and ControlModesControl ChannelControl channels command actuator movement by providing a valve driver signal to the ser

Strona 74

Sensor LimitsOne type of detector is a limit detector for sensor input signals. Each sensor can have a high and a lowlimit which you can enable separa

Strona 75

Why Tuning Is BeneficialTuning optimizes system response by adjusting the system so that its servoloop responds accurately toits command signal. When

Strona 77

Running the Example HCF TestTopics:•About This Chapter...

Strona 78 - Install the Specimen

DecommissioningDecommission the System...1245Ta

Strona 79

About This ChapterThis chapter describes how to set up and run the Example HCF Test using a typical MTS TestLine system.Adapting This Chapter to Your

Strona 80

The controller power switch is located on the back of the controller.Open the Station Manager Application1. Close any open applications on the compute

Strona 81

3. If necessary, select a project.It is common practice to keep all files in Project1. If you find this acceptable, at the Select Projectwindow, selec

Strona 82

Parameter sets are found in the lower left corner of the Open Station window. For this example HCFtest, select the HardSpecimenTuning parameter set an

Strona 83

a) Select Station Setup button> Channels > Ch1 (or other depending on your configuration) >Displacement > Channel Input Signals button >

Strona 84 - Prepare To Run Test

Verify That the System Is Ready for UseBefore operation, take a moment to familiarize yourself with the system while verifying that all componentsare

Strona 85

4. Verify that the actuator hydraulic pressure (P), return (R), and drain (D) lines are connected.5. Verify that the actuator force sensor (FS), servo

Strona 86

Set Initial Limit and Error DetectorsYou can reduce the chances of equipment damage or injury by setting limit and error detectors.Warning:Setting lim

Strona 87

a) Select Station Setup button > Channels > Ch1 > Displacement > Channel Input Signals Button> Limits tab.b) Set the Upper Limit to 25

Strona 88

a) Select Station Setup button > Channels > Ch1 > Force > Channel Input Signals Button > Limitstab.b) Set the Upper Limit to .05 kN.c)

Strona 89

6Table of Contents

Strona 90

These error detector limits will be used throughout setup, tuning, and testing.a) Click the Detectors button in the Station Manager main window.b) Sel

Strona 91

a) Open the Tuning: Ch1 Displacement window by selecting Station Setup button> Channels >Ch1 (or other depending on your configuration) > Dis

Strona 92

a) Similarly, open the Tuning: Ch1 Force tab by selecting Station Setup button > Channels > Ch1(or other depending on your configuration) > F

Strona 93

Leave the Operator Type selection set to Tuning because you will do more tuning once the specimenis installed.Power Up the Pump and HSMHPU setups vary

Strona 94

a) Select the Exclusive Control check box.b) For the HPU, click the Power Low, and then Power High buttons.c) For the HSM, click Power Low only.Verify

Strona 95

a) In the Station Manager main window, click the Manual Command button.b) Select the following:1. Channel — Ch1 (or other depending on your configurat

Strona 96

3. Verify polarity.a) Move the actuator about 10 mm in the positive direction. The actuator should extend.Note: If a displacement limit trips, see If

Strona 97

2. Set up a force waveform using the Function Generator.This introduces a force waveform to the specimen.a) In the Station Manager main window, click

Strona 98

Set up a Scope to compare the command signal to the resulting feedback. Enlarge the scope ifnecessary, and then set up the scope as follows:Y1 paramet

Strona 99

6. Run the Function Generator.Click the Program Run button to apply the force waveform to the specimen. If necessary, clear anyinterlock conditions.7.

Strona 100 - Check and Reset Meter History

Technical SupportHow to Get Technical SupportStart with your manualsThe manuals supplied by MTS provide most of the information you need to use and ma

Strona 101 - Set Limits for the Test

b) If P gain is insufficient, as shown, gradually increase P gain until the scope looks like that in step d.70 | MTS TestLine Test System OperationRun

Strona 102 - Minimize Station Manager

Insufficient P Gainc) If P gain is excessive and goes unstable, as shown, decrease your P gain by 20%.MTS TestLine Test System Operation | 71Running t

Strona 103 - Run the Example HCF Test

Excessive P Gaind) Adjust P gain until your scope looks similar to the illustration below.72 | MTS TestLine Test System OperationRunning the Example H

Strona 104

Correct P GainNote: The scope display shows how well the feedback signal is following the commandsignal. You should strive for a good match to the &qu

Strona 105 - Select the Test

9. Turn off the Function Generator.Click the Program Stop button to turn off the Function Generator.Position the Actuator Using Displacement ControlIn

Strona 106 - Create a New Test Run

Set the displacement limits to +40 mm and -40 mm so that you can achieve full extension and retraction.2. Use manual command to find the actuator trav

Strona 107

Retract the actuator fully, and then measure the length of the actuator so that you can later determinewhere to set your limit detectors.3. Position t

Strona 108 - Review the Results

Set HSM power to low, and then to off. If needed for other stations, the HPU can be left on.2. As an added precaution, set the handvalve (1) to off.MT

Strona 109 - Generate a Report

Install the SpecimenSet Up Meters for Specimen Installation1. Click the meters button in the Station Manager application.2. If more than one meter app

Strona 110

A running max/min meter provides the history of sensor extremes that the meter has read since it waslast reset.4. Set up a Running Max/Min Force Meter

Strona 111 - Remove the Specimen

When you have more than one MTS system, the system job number identifies your system. You can findyour job number in your order paperwork.Example syst

Strona 112 - Running the Example HCF Test

Use the following parameters:• Meter Type — Timed• Channel — Ch1 (or other depending on your configuration)• Signal — Force• Engineering Units — kNZer

Strona 113 - Displacement Limit

a) If necessary, click the Station Setup button.b) Select Channels > Ch1 > Force > Channel Input Signals button > Offset/Zero tab.c) Click

Strona 114

3. Connect the specimen to the load cell.4. Connect the angle bracket to the specimen.82 | MTS TestLine Test System OperationRunning the Example HCF T

Strona 115

5. Make sure that the system is properly aligned, and then tighten the specimen angle bracket to the testbed.6. Tighten the actuator angle bracket to

Strona 116

Prepare To Run TestSet Limits for TuningDuring tuning, it is possible for the system to go unstable. Setting limits just outside expected values fortu

Strona 117 - Maintenance

Because the specimen is fixed solidly in place, you can set narrow displacement limits that will trip ifanything breaks.a) Click the Station Setup but

Strona 118

The expected force range for tuning is 0.1 kN to 0.5 kN. However, the specimen is currently at zeroForce so that must be accommodated. For this exampl

Strona 119

Click the Reset button. If the interlock will not reset, there is likely an interlock condition that has notbeen corrected.3. Open the handvalve (1).4

Strona 120 - Maintenance Intervals

a) Select the Exclusive Control check box.b) For the HPU and HSM, click the Power Low, and then Power High buttons.Check for Force Control Drift Aroun

Strona 121 - Daily Inspections

a) Y1 parameters:• Channel — Ch1 (or other depending on your configuration)• Signal — Command• Unit — kN• Units/div — 0.2000• Offset — 0.0000b) Y2 par

Strona 122

— Messaging applicationsIf You Contact MTS by PhoneA Call Center agent registers your call before connecting you with a technical support specialist.

Strona 123 - Decommissioning

This example was not stable enough with the initial cold settings. As a result, the scope showed forcedecreasing to -0.6 kN which eventually tripped t

Strona 124 - Decommission the System

For this example, P gain was increased in increments of 1 until the drift around zero was corrected witha P gain of 4. I gain was then set to 25% of t

Strona 125

As you can see in the preceding illustration, the system is performing much better with the P gain setto 4.Note: Every system is different and require

Strona 126

This introduces a force waveform to the specimen.a) In the Station Manager main window, click the Function Generator button.b) Set up a force waveform

Strona 127

Set up a Scope to compare the command signal to the resulting feedback. Enlarge the scope ifnecessary, and then set up the scope as follows:Y1 paramet

Strona 128

6. Run the Function Generator.Click the Program Run button to apply the force waveform to the specimen. If necessary, clear anyinterlock conditions.7.

Strona 129

b) If P gain is insufficient, as shown, gradually increase P gain until the scope looks like that in step d.96 | MTS TestLine Test System OperationRun

Strona 130

Insufficient P Gainc) If P gain is excessive and goes unstable, as shown, decrease your P gain by 20%.MTS TestLine Test System Operation | 97Running t

Strona 131

Excessive P Gaind) Adjust P gain until your scope looks similar to the illustration below.98 | MTS TestLine Test System OperationRunning the Example H

Strona 132

Correct P GainNote: The scope display shows how well the feedback signal is following the commandsignal. You should strive for a good match to the &qu