SWIFT 20 Sensors Maintenance
123
Maintenance
Overview This chapter contains scheduling guidelines and detailed instructions for
performing preventive maintenance. Preventive maintenance is a set of routine
procedures that allow you to extend the operating life of your transducer and the
transducer interface electronics. You can prevent excessive wear or possible
component failure through regular inspections and simple procedures, such as
filter cleaning.
The information provided in this chapter is a recommendation only. The actual
time intervals will depend on the operating conditions at your facility.
Contents Transducer 124
Low-Profile Transducer Interface 125
Cables 126
Maintenance Schedule
Activity Customer Preventive Maintenance Contact MTS
Calendar Time 1 Day 1 Week 1 Month 1 Year Suggested
Inspect Transducer for cracks X
Check Transducer Interface cooling X
Clean Transducer Interface fan filter X
Inspect electrical cables X
Calibrate the Transducer X