be certain.mProduct InformationModel 680.01 High-Temperature Grips015-018-801 F
Model 680.01 High-Temperature Grips10ConventionsConventionsDocumentation ConventionsThe following paragraphs describe some of the conventions that are
ConventionsModel 680.01 High-Temperature Grips 11Hypertext links The electronic document has many hypertext links displayed in a blue font. All blue w
Model 680.01 High-Temperature Grips12Conventions
Model 680.01 High-Temperature Grips 13IntroductionThe MTS Model 680.01 High-Temperature Hydraulic Grips are used with a high temperature furnace insta
Model 680.01 High-Temperature Grips14Model 680.01 High-Temperature Grips Functional DescriptionThe pushrod, grip extension, and replaceable specimen a
Model 680.01 High-Temperature Grips 15Model 680.01 High-Temperature Grip AccessoriesThe following accessories are available for the grips:• Specimen a
Model 680.01 High-Temperature Grips16mount a grip to the actuator rod. Attachment kits are not supplied with a grip set, and they must be ordered sepa
Model 680.01 High-Temperature Grips 17Hand-operatedhydraulic pumpA hand-operated hydraulic pump kit (MTS part number 038-634-701) is a required access
Model 680.01 High-Temperature Grips18Model 680.01 High-Temperature Grips DimensionsMounting holes in the base of each grip are threaded in either SI I
Model 680.01 High-Temperature Grips 19Safety InformationHazard Placard PlacementHazard placards contain specific safety information and are affixed di
Copyright information © 1991, 2001, 2002, 2008, 2009 MTS Systems Corporation. All rights reserved.Trademark information MTS is a registered trademark
Model 680.01 High-Temperature Grips20
Model 680.01 High-Temperature Grips 21InstallationThis section describes how to install the Model 680.01 High Temperature Grips into a load unit.Conte
Model 680.01 High-Temperature Grips22Install the GripsAn attachment kit is required for attaching each grip to the load unit. Kit selection should be
Model 680.01 High-Temperature Grips 23Preparation Ensure that the threads of all threaded components in the force train (between the load cell and the
Model 680.01 High-Temperature Grips24B. With the connector stud is facing up, add any required shims, spacers, or spiral washers to the stud. Ensure t
Model 680.01 High-Temperature Grips 25Install the Hand-Operated Hydraulic PumpWhen space permits, the hand-operated hydraulic pump can be mounted to t
Model 680.01 High-Temperature Grips26Install the Cooling KitCooling kits circulate water through the grips to cool the them when they are used in high
Model 680.01 High-Temperature Grips 27Preload the Spiral WashersThe components in the force train must be preloaded to prevent backlash during operati
Model 680.01 High-Temperature Grips286. Using spanner wrenches, rotate the two halves of the spiral washers in opposite directions to tighten them. Th
Model 680.01 High-Temperature Grips 29Grip Water Cooling Assembly Configurations for Servohydraulic FramesCut hoses to required length at installation
Model 680.01 High-Temperature Grips Contents3ContentsTechnical Support 5How to Get Technical Support 5Before You Contact MTS 5If You Contact
Model 680.01 High-Temperature Grips30048-974-804Hoses towaterservice
Model 680.01 High-Temperature Grips 31048-974-805Hoses towaterservice
Model 680.01 High-Temperature Grips32
Model 680.01 High-Temperature Grips 33OperationThis section describes how to use the grips.Contents Determine the Gripping Pressure 33Install Threa
Model 680.01 High-Temperature Grips34Grip Force Rating versus Specimen TemperatureForce versus hydraulicpressureThe horizontal scale at the bottom of
Model 680.01 High-Temperature Grips 35Install Threaded Specimens Perform the following procedure to mount threaded specimens in the grips. Ensure that
Model 680.01 High-Temperature Grips36Hydraulic power will be applied to the systems hydraulic actuator during the following steps.This can cause perso
Model 680.01 High-Temperature Grips 37Install Button-End Specimens Perform the following procedure to mount button-end specimens in the grips. Ensure
Model 680.01 High-Temperature Grips38Hydraulic power will be applied to the systems hydraulic actuator during the following steps.This can cause perso
Model 680.01 High-Temperature Grips 39Remove Specimens from the GripsParts of the grips can be very hot after gripping specimens inside a high-tempera
Model 680.01 High-Temperature Grips4ContentsOperation 33Determine the Gripping Pressure 33Install Threaded Specimens 35Install Button-End Spe
Model 680.01 High-Temperature Grips40
Model 680.01 High-Temperature Grips 41MaintenanceThe Model 680.01 High-Temperature Grips are designed to be used in high temperature environments. Af
Model 680.01 High-Temperature Grips42Add Hydraulic Fluid to the Hand PumpCheck the fluid level in the reservoir of the hand pump occasionally by remov
mMTS Systems Corporation14000 Technology DriveEden Prairie, Minnesota 55344-2290 USAToll Free Phone: 800-328-2255 (within the U.S. or Canada)Phone: 95
Model 680.01 High-Temperature Grips 5Technical SupportHow to Get Technical SupportStart with yourmanualsThe manuals supplied by MTS provide most of th
Model 680.01 High-Temperature Grips6Know information fromprior technicalassistanceIf you have contacted MTS about this problem before, we can recall y
Model 680.01 High-Temperature Grips 7If you are calling about an issue that has already been assigned a notification number, please provide that numbe
Model 680.01 High-Temperature Grips8
Model 680.01 High-Temperature Grips 9PrefaceBefore You BeginSafety first! Before you attempt to use your MTS product or system, read and understand th
Komentarze do niniejszej Instrukcji