  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 72
5 +
Years
2 Year1 Year6
months
3
months
2
months
1
Month
1
Week
DailyCalendar time
10,0004,0002,0001000500320160408Running time - Hours
MTSXCheck functionality of
warning and interlock
devices
XConduct external
inspection of the heat
exchanger
XReplace fluid filter
MTSInspect incoming power
lines for loose
connections or defects
XCheck auto-cooling
circuit functionality
XCheck water hoses
MTSReplace PLC battery
Controller
XCheck cooling
XPerform shunt
calibration
MTSVerify controller
calibration
Verify Safety Circuits
XCheck E-Stops
XCheck limits
XCheck test area guard
interlocks
XCheck actuator speed
control (370 Load
Frame)
Environmental
Components (optional)
MTS Material Test System Operation 73
8.0 Routine Maintenance